Working Principle
Through the oil-free air compressor which set in the control box to fill the fresh and clean air in cabin,and form a micro pressure environment of 1.6 atmospheres. And adopt medical grade oxygen generator to provide the oxygen for the cabin personnel through the oxygen supply pipeline.
Technical data
1.Design pressure: 0.07MPa, maximum working pressure: 0.06Mpa
2.Seating capacity: one person
3.Pressurized medium: air
4.Cabin volume: ≥2.5 m3
5. Oxygen concentration in Cabin: ≤ 23%, when ≥ 23%, there will be sound and light alarm
6. Relative humidity: ≤ 80%
7. Pressurization time for the chamber: ≤ 15min
8. Power supply: 220V 50Hz (or customized)
9. Working environment temperature: 0 ℃~ + 30 ℃
10. Noise in chamber: ≤ 65dB
11. Control mode
Automatic control: PLC automatic control
Manual control: manual control valve inside and outside cabin (emergency)
12. Size: L2066*W1235*H2160mm
13. Weight: about 1300kg