Hard shell lying oxygen chamber

Hard shell lying oxygen chamber

Working Principle

Through the oil-free air compressor set in the control box, the cabin is filled with fresh and clean air to form a micro pressure environment of 1.5 atmospheres. The medical grade oxygen generator or oxygen bottle is used to provide the oxygen needed for oxygen inhalation to the cabin personnel through the oxygen supply pipeline.

Technical Index

1. Envelope dimension of cabin : Φ 800 mm * 2450 mm

2. Maximum working pressure : 0.05 MPa

3. Rated number of passengers entering the cabin : 1 person 

4. Cabin leakage rate ≤ 5 % / h

5. Pressure reducing noise ≤ 65 dB ( A )

6. Pressurized medium : air 

7. Oxygen concentration in cabin ≤ 23 %

8. Empty weight of cabin ≤ 300 Kg

9. Volume : 1.2 m³

10. Normal working time of equipment : ≤ 10 min ( up to maximum working pressure ) 

11. Power Supply : AC 220 V ± 33 V , 50 HZ ± 2 HZ or customized

12. Working environment temperature : 0 ~ 30 ℃

13. Storage limit temperature :  -10 ~ 50 ℃

14. Relative humidity : ≤ 95 %